Endovascular and catheter technologies designed for exacting control and minimally invasive precision.
Radiofrequency energy is delivered from a console that does not simply output a fixed wattage. The catheter carries a temperature sensor at the heating element and reports back continuously; the generator's algorithm then raises or lowers power to hold the set target. Impedance is monitored on the same circuit.
That closed loop is the difference between a dose and a setting. If the segment is not fully exsanguinated, or the catheter is not apposed to the wall, the console sees it as a temperature shortfall at high power and says so — rather than quietly delivering an inadequate treatment.
A sterile, single-use bipolar catheter. No return electrode and no patient plate: the circuit is contained within the element itself, which is what allows the working length to be treated segment by segment rather than continuously.
The heating element is echogenic, so tip position is confirmed on duplex rather than inferred. Shaft markers set the pull-back increment — when the last treatment mark clears the sheath hub, the run is complete.
The procedure runs under tumescent local anaesthesia through a percutaneous puncture. There is no groin incision, no general anaesthetic and no theatre admission — which is what makes it viable in an office theatre rather than a hospital list.
Tumescent infiltration does two jobs at once: it anaesthetises the perivenous space and it puts a fluid layer between the vein and the skin, which is the principal defence against thermal injury to skin and adjacent nerves.
Not supplied with the system:
Treatment cycle
The generator splits each cycle into a heating phase and a hold phase. Default settings suit most anatomy; both phases are adjustable from the console.
Tip placed 2 cm below the SFJ or SPJ under duplex, tumescent infiltrated circumferentially.
Confirmed on ultrasoundPower ramps to bring the element to target. The console counts down to temperature.
6 s · up to 40 W · 120 °CPower drops back and the algorithm holds the target while collagen contracts.
14 s · 25 W · 120 °CPull back to the next shaft marker and repeat. Output stops automatically at the end of each cycle.
Segmental, marker to marker| Element | Phase 1 time | Phase 1 power | Phase 2 time | Phase 2 power | Target |
|---|---|---|---|---|---|
| 7 cm | 6 s | 40 W | 14 s | 25 W | 120 °C |
| 5 cm | 6 s | 30 W | 14 s | 18 W | 120 °C |
| 3 cm | 6 s | 18 W | 14 s | 11 W | 120 °C |
| 2 cm | 6 s | 12 W | 14 s | 7 W | 120 °C |
| 1 cm | 6 s | 6 W | 14 s | 4 W | 120 °C |
Information on this page is intended for healthcare professionals. Full indications, contraindications, warnings and potential complications are set out in the manufacturer's Instructions for Use, which take precedence over any summary presented here.