Engineering Precision Flow

Endovascular and catheter technologies designed for exacting control and minimally invasive precision.

Endovascular Systems

Radiofrequency energy is delivered from a console that does not simply output a fixed wattage. The catheter carries a temperature sensor at the heating element and reports back continuously; the generator's algorithm then raises or lowers power to hold the set target. Impedance is monitored on the same circuit.

That closed loop is the difference between a dose and a setting. If the segment is not fully exsanguinated, or the catheter is not apposed to the wall, the console sees it as a temperature shortfall at high power and says so — rather than quietly delivering an inadequate treatment.

What the console monitors

  • Element temperature, 10–130 °C, accurate to ±3 °C
  • Delivered RF power against the catheter's rated ceiling
  • Circuit impedance, with alarms outside 32–300 Ω
  • Cycle countdown, cumulative treatment time and cycle count

Automatic protection

  • RF output is impossible unless a recognised catheter is connected
  • Output cuts automatically above 130 °C at the element
  • Output cuts automatically when the set cycle time expires
  • Below 25 °C at start, the console warns that the catheter may not be in the vein
  • Sustained low temperature at high power prompts a compression check
  • Self-calibration runs on every power-up; no scheduled service interval

Catheter Technologies

A sterile, single-use bipolar catheter. No return electrode and no patient plate: the circuit is contained within the element itself, which is what allows the working length to be treated segment by segment rather than continuously.

The heating element is echogenic, so tip position is confirmed on duplex rather than inferred. Shaft markers set the pull-back increment — when the last treatment mark clears the sheath hub, the run is complete.

Choosing between the families

  • CFB 2.3 — the routine choice for great saphenous work
  • CFA 2.3 — adds a 0.025″ guidewire lumen for tortuous segments
  • CFC 2.0 and CFD 1.6 — narrower shafts on a 6F sheath, for smaller-calibre veins
  • Short elements (1–3 cm) — below-knee segments, accessory veins and short refluxing lengths
  • 100 cm working length — for longer limbs
Heating element Position markers Ball-end tip 480 kHz · 95–120 °C · bipolar

Minimally Invasive Solutions

The procedure runs under tumescent local anaesthesia through a percutaneous puncture. There is no groin incision, no general anaesthetic and no theatre admission — which is what makes it viable in an office theatre rather than a hospital list.

Tumescent infiltration does two jobs at once: it anaesthetises the perivenous space and it puts a fluid layer between the vein and the skin, which is the principal defence against thermal injury to skin and adjacent nerves.

Volume guide. Approximately 10 cc per cm of vein to be treated, infiltrated circumferentially, with more than 1 cm of separation created between the anterior vein wall and the skin wherever the vein runs superficially.

Also required in theatre

Not supplied with the system:

  • Tilt table
  • Duplex ultrasound scanner and probes
  • Sterile ultrasound couplant
  • Puncture needle
  • Introducer sheath, 6F or 7F, 7 cm or 11 cm
  • 0.025″ guidewire, where a CFA catheter is used

Treatment cycle

Two phases, one pull-back

The generator splits each cycle into a heating phase and a hold phase. Default settings suit most anatomy; both phases are adjustable from the console.

Position

Tip placed 2 cm below the SFJ or SPJ under duplex, tumescent infiltrated circumferentially.

Confirmed on ultrasound

Heating phase

Power ramps to bring the element to target. The console counts down to temperature.

6 s · up to 40 W · 120 °C

Hold phase

Power drops back and the algorithm holds the target while collagen contracts.

14 s · 25 W · 120 °C

Withdraw

Pull back to the next shaft marker and repeat. Output stops automatically at the end of each cycle.

Segmental, marker to marker
Factory default treatment settings, by heating element
ElementPhase 1 timePhase 1 power Phase 2 timePhase 2 powerTarget
7 cm6 s40 W14 s25 W120 °C
5 cm6 s30 W14 s18 W120 °C
3 cm6 s18 W14 s11 W120 °C
2 cm6 s12 W14 s7 W120 °C
1 cm6 s6 W14 s4 W120 °C
Limit. No venous segment should be treated for more than three cycles — ablation depth does not increase meaningfully beyond that, and the catheter should not be reintroduced into a segment already treated.